ROUND UP RU 200

A sealed mixing system for autologous, heterologous or synthetic graft materials.

What it treats

Applications

  • Orthopaedic use, which DNA Nexus records for this device and Silfradent does not name on its own ROUND UP page.
  • Liquid mass homogenization.
  • Semiliquid homogenization.
  • Semisolid homogenization.
  • Integrated homogenization.
The mechanism

How it works

ROUND UP works by intrinsic and extrinsic molecular blending: it mixes autologous, heterologous or synthetic graft materials without altering their geometric dimensions. The chamber is sealed, which Silfradent publishes as mixing free from atmospheric contamination, and the operator runs one of four preset programmes before the system empties itself automatically, in sixteen seconds at most.

Liquid mass homogenization

3 to 5 seconds

Semiliquid homogenization

6 to 8 seconds

Semisolid homogenization

9 to 12 seconds

Integrated homogenization

12 to 16 seconds

Automatic empty

Up to 16 seconds, once mixing ends

In the room

Why this system

Sealed mixing, three graft-material categories

ROUND UP handles autologous, heterologous or synthetic graft materials in one sealed system.

Part of the Silfradent range built on CGF, biofiller and sticky bone

Silfradent's own positioning names CGF, the APAG biofiller and sticky-bone preparation. ROUND UP is the mixer DNA Nexus carries from that range, alongside MEDIFUGE CGF and APAG.

One system for three material states

ROUND UP runs liquid, semi-liquid or solid materials through the same device.

A Class IIa medical device

ROUND UP carries the CE Class IIa classification.

The detail

What to know

Practice value

  • Class: IIa
  • Speed: 3,600 rpm at 60 Hz
  • Timer: 1 to 16 seconds
  • Decontamination: UVC cycle, 5 minutes
  • Mains: 230V or 100 to 115V, 320 VA
  • Dimensions: 32 x 32 x 30 cm
  • Weight: 19 kg

What materials does ROUND UP mix?

Autologous, heterologous or synthetic graft materials, in a sealed system.

How fast does ROUND UP run?

3,600 rpm at 60 Hz, on a timer of 1 to 16 seconds. Silfradent publishes the speed at that frequency only, and lists no equivalent figure for 50 Hz mains.

How is ROUND UP decontaminated between uses?

A UVC cycle, 5 minutes.

Does ROUND UP connect to Silfradent's CGF protocol?

Silfradent's positioning names CGF, the APAG biofiller and sticky-bone preparation as what its range is built on. ROUND UP is the mixer DNA Nexus carries from that range, alongside MEDIFUGE CGF and APAG. Silfradent names no procedure that relies on ROUND UP specifically.

Is a variant available for a different electrical standard?

Yes. Alongside the standard RU 200 (230V, 50/60 Hz), Silfradent lists the RU 200 100 as an optional variant, wired for 100 to 115V, 50/60 Hz.

How fast does ROUND UP empty once mixing is done?

Automatically, in a few seconds, 16 seconds at most.

DNA Nexus support

DNA Nexus will handle installation and set up ROUND UP RU 200 on site.

ROUND UP RU 200 - Silfradent